Design and optimization of an alternative high efficiency sample introduction system for single particle ICP-MS analysis
In this study we present the development process of an alternative sample introduction system for inductively coupled plasma mass spectrometers (ICP-MS) with the aim of improving the efficiency and capabilities of single particle ICP-MS measurements. The system consists of a high efficiency pneumati...
Elmentve itt :
Szerzők: | |
---|---|
Testületi szerző: | |
Dokumentumtípus: | Könyv része |
Megjelent: |
2022
|
Sorozat: | Proceedings of the International Symposium on Analytical and Environmental Problems
28 |
Kulcsszavak: | Analitikai kémia |
Tárgyszavak: | |
Online Access: | http://acta.bibl.u-szeged.hu/78506 |
Tartalmi kivonat: | In this study we present the development process of an alternative sample introduction system for inductively coupled plasma mass spectrometers (ICP-MS) with the aim of improving the efficiency and capabilities of single particle ICP-MS measurements. The system consists of a high efficiency pneumatic nebulizer, and a low memory effect, high transport efficiency onaxisspray chamber which are currently under manufacturing utilizing a high resolution, multijet (MJP) 3D printer. Both part underwent several fluid dynamic simulations to debunk potential design flaws and to assess their optimal operating conditions before their production. |
---|---|
Terjedelem/Fizikai jellemzők: | 187-190 |
ISBN: | 978-963-306-904-2 |